Re: Spectrophotometer Help

From: David Cross (spamdenied_at_nospam.net)
Date: 03/27/05


Date: Sun, 27 Mar 2005 14:21:46 -0800

On 27 Mar 2005 13:36:25 -0800, "Amemoto" <hateame@yahoo.com> wrote:

>wavelength(nm) absorbance
>340 0.706
>360 1.070
>380 error(-1.999)
>400 error(-1.999)
>480 error(-1.999)
>500 1.900
>520 1.230
>
>I don't know whether the problem is with the old instrument or not, but
>i suspect it has to do with the concentration of my solution. ANY
>IDEA???Has this encountered anyone before?Please let me know what the
>problem is. tia

Yeah, sounds like your machine was registering out of range due to your
solution being too concentrated.

---
David Cross
dcross1 AT shaw DOT ca


Relevant Pages

  • Re: Labview VI behaving inconsistently
    ... I'm not certain how much info I can give you.  I suspect the C ... The purpose of my query was to see if this might be a common problem ... the instrument control VIs are homemade and have ...
    (comp.lang.labview)
  • Re: Survey: Favorite single recording of each Beethoven symphony?
    ... > Lena wrote: ... I suspect I know nothing about the instrument. ... admit this whole issue is not one I'm constantly all that concerned ...
    (rec.music.classical.recordings)
  • Re: RIP Oscar Peterson
    ... A Jazz musician. ... I don't recall his instrument, but I think it is (or I ... suspect was due to the tense of the original post) the piano. ...
    (uk.radio.amateur)
  • Re: Live 02 Well done!
    ... bands that would be a v good thing. ... But i suspect we'd get second rate synth pop shite... ... Or we would have got sugababes, or some other non entity singing outfit with a backing track who wouldn't know an instrument if they tripped over it on the way to the dressing room. ...
    (uk.rec.scouting)