The answer depends on the kind of signal you need to measure and type of
oscilloscope used.

A typical oscilloscope vertical amplifier input will have a 1M ohm input
impedance. The typical 10x passive probe will have a 10M ohm impedance.

For signals of 350MHz or less conventional passive probes are fine.

Oscilloscope designed for signals from 500MHz to 4000MHz will benefit
from using active probes.

Active probes have high bandwidth amplifiers at or near the probe tip.

The amplifiers output is designed to drive a 50 ohm cable to the
oscilloscope vertical amplifier input.

Oscilloscope that are intended to use only active probes will usually
have only a 50 ohm input impedance.

-----------

The basic concept here is that it takes a lot of power to accurately
drive a 500MHz signal down a 1 meter cable. Most circuits that use
signals at this frequency do not have the needed power and are adversely
affected when connected to the load of an oscilloscope vertical
amplifier input.

The ideal oscilloscope probe would have very high impedance, low
capacitance and disturb the signal very little.

As the signal frequency increases it becomes much harder to make a probe
that can do what is needed.



Relevant Pages

  • Re: Advice on HP Function Generator Repair
    ... using an external oscilloscope to look at the FG's output. ... the oscilloscope probe and/or the oscilloscope itself not ... able to handle incoming frequencies that high? ... If it is definitely coming from the function generator, ...
    (sci.electronics.repair)
  • Re: Advice on HP Function Generator Repair
    ... using an external oscilloscope to look at the FG's output. ... the oscilloscope probe and/or the oscilloscope itself not ... you glue a low-pass filter on the end of such a generator. ... place the termination at the scope input rather than at the ...
    (sci.electronics.repair)
  • Re: Advice on HP Function Generator Repair
    ... using an external oscilloscope to look at the FG's output. ... the oscilloscope probe and/or the oscilloscope itself not being ... able to handle incoming frequencies that high? ... If it is definitely coming from the function generator, ...
    (sci.electronics.repair)
  • Re: Enigmatic Frequency: 400 Hz
    ... > Have you ever experienced the oscilloscope shows the frequency around ... > 400 Hz when you grasp the probe of that oscilloscope with keeping the ... I can make a better diagnosis of your mental health by reading your posts. ...
    (sci.med)
  • Re: Linux, PC-based, PCI or USB oscilloscopes?
    ... Especially for a beginner I recommend a analog oscilloscope, ... purposes on _slow_ signals. ... sampling rate of at least 2*n, which is described by the Nyquist ...
    (comp.os.linux.hardware)